1. Home
  2. testo Saveris H3

testo Saveris H3 - 2-channel temperature/humidity radio probe

0572 6250
  • Capacitive humidity sensor with long-term stability
  • Large measurement data memory
  • Flexible wireless data transfer
  • Recording of temperature and humidity
  • The testo Saveris H3 radio probe is equipped with an internal humidity and temperature probe and is therefore ideal for monitoring warehouse goods that are sensitive to temperature and humidity. It also monitors the air conditioning in buildings, continuously and reliably.

    The testo Saveris H3 radio probe is equipped with an internal humidity and temperature probe and is therefore ideal for monitoring warehouse goods that are sensitive to temperature and humidity. It also monitors the air conditioning in buildings, continuously and reliably.

    The testo Saveris H3 guarantees reliable measurement results based on the latest
    measuring technology. The integrated sensors guarantee readings with long-term stability. This enables quality assurance guidelines to be complied with and documented securely. The testo Saveris H3 radio probe has a data memory for 6,000 readings per channel.

    Please note: Saveris converter V 2.0 (order no. 0572 0258) is required for integrating Saveris radio probes into systems with base firmware V 1.X. For more information please contact your Testo partner.

    testo Saveris H3 humidity/temperature radio probe, radio frequency 2.4 GHz, including wall bracket, calibration protocol and 4 × AA alkaline manganese batteries.

    General technical data

    Weight
    245 g
    Dimensions
    80 x 85 x 38 mm
    Operating temperature
    -20 to +50 °C
    Protection class
    IP42
    Product-/housing material
    Plastic
    Radio frequency
    2.4 MHz
    Transmission distance
    approx. 100 m without obstruction at a frequency of 2.4 GHz
    Wall bracket
    included
    Measuring rate
    Standard 15 min, 1 min to 24 h can be set
    Battery type
    4 x type AA batteries
    Battery life
    Battery life at +25 °C, 3 years; for freezer applications, 3 years with L91 Photo lithium Energizer batteries
    Memory
    6,000 measurement values / channel
    Storage temperature
    -40 to +55 °C (incl. batteries)