| Operating humidity | 10 to 90 %RH non-dewing |
|---|---|
| Operating temperature | 5 to +45 °C |
| Protection class | IP40 |
| Storage temperature | -20 to +50 °C |
| Measuring range | -200 to +300 °C (with surface probe 0604 0994 (accessory)) Depends on probe, max. -40 to +1200 °C |
|---|---|
| Accuracy | ±0.5 % of mv (Remaining Range) additionally ±1 digit, additionally accuracy of probe ±0.5 °C (-40 to +100 °C) |
| Resolution | 0.1 °C |
| Measuring rate | max. 1 / s |
| Measuring range | -9999.9 to +9999.9 Pa |
|---|---|
| Accuracy | ±0.3 Pa (0 to +9.99 Pa) ±3 % of mv (+10.00 to +9999.9 Pa) additionally ±1 digit |
| Resolution | 0.01 Pa (-149.99 to +149.99 Pa) 0.1 Pa (Remaining Range) |
| Measuring rate | max. 1 / s |