The greatest measurement errors in thermography are almost always caused by three variables: incorrectly set emissivity (ε), unconsidered reflected temperature (RTC), and a measurement distance at which the target is imaged too small in the thermal image (IFOV). If you consistently check and document these three points, thermograms become much more reliable – especially for metals, small components and changing environments.
First rule: a thermal imaging camera always shows the temperature of the visible surface – not the temperature under a cover or layer of dirt. For reliable temperature values, emissivity (ε) and reflected temperature (RTC) must match the surface and the measurement environment.
The colour of a surface has little practical influence on long-wave infrared radiation; temperature and surface condition are decisive. Smooth, glossy, reflective or polished surfaces usually have lower emissivity than matt, rough or weathered surfaces of the same material – and therefore reflect interference sources from the surroundings more strongly.
A thermography measurement physically consists of three components: emission (ε) from the measured object, reflection (ρ) of ambient radiation and, where applicable, transmission (τ) through certain materials. In practice, τ is almost 0 for most solid materials – exceptions include thin plastic films, for example.
For many surfaces, approximately: ρ = 1 − ε. Low ε means a high proportion of reflection – and therefore greater sensitivity to interference sources and incorrect RTC compensation.
The direction of the measurement error depends on whether the measured object is warmer or colder than the surroundings.
ε set too high → temperature display too low
ε set too low → temperature display too high
ε set too high → temperature display too high
ε set too low → temperature display too low
Remember: the greater the temperature difference between the measured object and the surroundings and the lower ε is, the greater the measurement errors become – and they are amplified when ε is set incorrectly.
Emissivity tables provide reference values: ε changes with temperature and surface condition, among other factors. Use table values as a starting point and, if necessary, validate them using a reference surface/reference method before documenting absolute temperature values.
Examples of typical reference values:
Concrete: ε ≈ 0,93
Glass (warm): ε ≈ 0,9
Aluminium (not oxidized): ε ≈ 0,02–0,03 (highly reflective → RTC/interference sources particularly critical)
Set emissivity (ε) to match the material/surface,
take reflected temperature (RTC) into account when surfaces are reflective or the environment has strong temperature contrasts,
choose the measurement distance so that the target is imaged clearly larger than one pixel (IFOV).
In addition, focus and consistent scale logic ensure comparability.
Emissivity describes how effectively a surface emits IR radiation. Low ε means that a large proportion of the “seen” radiation is reflected from the surroundings. This can severely distort absolute temperature values – even if the thermal image looks “plausible”.
ε is between 0 and 1. High ε (e.g. many matt, non-metallic surfaces) is “forgiving” in thermography. Low ε (typical of bare metal) is critical because reflections dominate.
Measured values change significantly with viewing angle or location.
“Hotspots” appear, disappear or move.
The camera apparently shows the temperature of people, the sky or warm objects in the surroundings because they are reflected.
Practical measures include creating a suitable reference surface (e.g. a defined, well-emitting point), documenting measurement conditions (distance, viewing angle) and planning a comparative measurement (contact/another method) at critical points.
If you work with emissivity tables, use the values as a starting point and validate the result using a reference surface and reproducible measurement conditions. This prevents a “table value” from being mistaken for false precision.
RTC is the effective ambient temperature that enters the measured value as a reflection. RTC is particularly relevant for reflective surfaces and when the surroundings are not “neutral” (e.g. warm machines, radiators, people nearby).
RTC is particularly important when:
the measured object is glossy/reflective,
the surroundings have strong temperature contrasts,
you use absolute temperatures or limit-value assessments.
In practice, the following help: shading/shielding against interfering radiation sources, a viewing angle that is as constant as possible and a documented measurement position. The goal is not “perfection”, but reproducibility.
IFOV answers the question: is my target measurable at all from this distance with this optics? If a target is only one or a few pixels in size, the pixel inevitably contains background components – the measured value is a mixed temperature.
A pixel is not a point, but represents an area/an angular range in the image. If the object is not imaged large enough, its informative value decreases. Small electrical connections, thin pipes or PV connections are particularly affected.
As a practical rule: the target should be clearly larger than a single pixel. If this is not possible, reduce the distance or choose optics/camera resolution that images the target larger.
Define the measurement objective (object, distance, purpose: screening vs. absolute temperature).
Check and fix the focus.
Set ε appropriately (material/surface).
Check RTC (reflective scene? strong temperature sources in the surroundings?).
Check distance/IFOV: is the target large enough in the image?
Choose the scale mode consciously (fixed/auto) and keep it consistent for comparative measurements.
Document: distance, viewing angle, scale values, measurement points.
Error pattern: the terminal appears “hot”, but the hotspot shifts when the location changes. Interpretation: reflection dominates (low ε) and/or the target is too small (IFOV). Countermeasures: prepare the measurement location as a reference surface, stabilize the viewing angle, reduce the distance or use telephoto optics, check the focus and document measurement points consistently. Then verify the cause using suitable electrical measurements.
Once ε/RTC/IFOV are understood as sources of error, camera selection can be better justified: resolution/optics suitable for the distance, NETD suitable for the contrast and workflow functions for documentation.