testo 330-2 LX - Industrial combustion analyzer kit 1

Order-Nr.  0563 3372 71

  • Long life sensors (up to six years) with over-range protection

  • Industrial CO range of 8,000 ppm (up to 30,000 ppm via onboard dilution)

  • Onboard instrument diagnostics and system leak check

  • Optional NOx sensor is field upgradable

The testo 330-2 LX combustion analyzer measures O2, CO (up to 30,000 ppm), temperature, pressure, and other required combustion efficiency parameters.  In addition, this flue gas analyzer has NOx upgradability, on-site printout capability, a large memory, and a graphic tuning chart to show ideal combustion efficiency.

Product Description

For commercial and industrial flue gas applications, the testo 330-2 LX combustion analyzer is properly outfitted with high range CO, sensor protection, a large memory for maintaining service records, on-site printing capability, and if needed, is easily field upgraded with a plug and play NOx sensor.  The interface for burner control modules makes this a versatile flue gas tuning and testing tool.  The testo 330-2 LX combustion analyzer has the longest sensor warranty on the market – 4 years!  Rugged construction and high quality sensors provide years of combustion efficiency, reliability, and a very low cost of ownership.
 
  • Built-on dilution systems allow for CO readings of up to 30,000 ppm
  • Onboard self-diagnostics and service tracking
  • Over-range sensor protection
  • Efficient "in the green" or "in the red" display box results allow for fine tuning combustion processes
  • Optional probes for high temperature, gas leak, ambient CO, and more
  • Large memory for up to 500,000 readings
  • Calculates CO2, efficiency, excess air, CO air free
 
For testing solid fuel during start up and shut down, use solid fuel adapter set 0600 9765.

 

Delivery Scope

testo 330-2 LX combustion analyzer, 12" flue gas probe with 7’ hose, extra particulate filters (10), AC power supply, rechargeable battery, certificate of conformity, and case
Differential Pressure

Measuring range

±40 InH₂O / ±10000 Pa

Accuracy

±0.0 InH₂O (0 to 0.04 InH₂O) plus ±1 Digit / ±0.3 Pa (0 to 9.99 Pa) plus ±1 Digit

±0 InH₂O (0 to 40 InH₂O) plus ±1 Digit / ±3 Pa (10 to 10000 Pa) plus ±1 Digit

Flue gas O₂

Measuring range

0 to 21 Vol.%

Accuracy

±0.2 Vol.%

Resolution

0.1 Vol.%

Reaction time t₉₀

< 20 sec

Flue gas CO

Measuring range

0 to 8000 ppm

Accuracy

±10 ppm or ±10 % of mv (0 to 200 ppm)

±20 ppm or ±5 % of mv (201 to 2000 ppm)

±10 % of mv (2001 to 8000 ppm)

Resolution

1 ppm

Reaction time t₉₀

< 60 sec

CO determination (with H₂-compensation), automatic dilution

Measuring range

0 to 30000 ppm

Accuracy

±100 ppm (0 to 1000 ppm)

±10 % of mv (1001 to 30000 ppm)

Resolution

1 ppm

Flue Gas NO

Measuring range

0 to 3000 ppm

Accuracy

±5 ppm (0 to 100 ppm)

±5 %rH (101 to 2000 ppm)

±10 % of mv (2001 to 3000 ppm)

Resolution

1 ppm

Reaction time t₉₀

< 30 sec

Flue gas NOlow

Measuring range

0 to 300 ppm

Accuracy

±2 ppm (0 to 39.9 ppm)

±5 % of mv (40 to 300 ppm)

Resolution

0.1 ppm

Reaction time t₉₀

< 30 sec

Flue Gas Draft

Measuring range

-4.00 to +16 InH₂O / -9.99 to +40 hPa

Accuracy

±0.01 InH₂O or ±5 % of mv (-0.20 to +0.24 InH₂O) / ±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa)

±0.01 InH₂O (+0.24 to +1.20 InH₂O) / ±0.03 hPa (+0.61 to +3.00 hPa)

±1.5 % of mv (+1.20 to +16.00 InH₂O) / ±1.5 % of mv (+3.01 to +40.00 hPa)

Resolution

0.00 InH₂O / 0.01 hPa

Temperature

Measuring range

-40° to 2192 °F / -40 to +1200 °C

Accuracy

±1 °F (32° to 212.0 °F) / ±0.5 °C (0 to +100.0 °C)

±0.5 % of mv (Remaining Range)

Resolution

0.1 °F (-40° to 1831.8 °F) / 0.1 °C (-40 to +999.9 °C)

1.0 °F (> +1832 °F) / 1 °C (> +1000 °C)

Efficiency (calculated)

Measuring range

0 to 120 %

Resolution

0.1 %

Flue Gas loss (calculated)

Measuring range

0 to 99.9 %

Resolution

0.1 %

Flue Gas CO₂ calculation (calculated from O₂)

Measuring range

0 to CO₂ max (Display range)

Accuracy

±0.2 Vol.%

Resolution

0.1 Vol.%

Reaction time t₉₀

< 40 sec

Velocity / Volume flow

Measuring range

29.50 to 591 fpm / 0.15 to 3 m/s

Resolution

0.1 fpm / 0.1 m/s

Pressure measurement

Measuring range

0 to +120 InH₂O / 0 to +300 hPa

Accuracy

±0.2 InH₂O (0.0 to 20.0 InH₂O) / ±0.5 hPa (0.0 to 50.0 hPa)

±1 % of mv (20.0 to 40.0 InH₂O) / ±1 % of mv (50.1 to 100.0 hPa)

±1.5 % of mv (Remaining Range)

Resolution

1{#0.0N} InH₂O / 0.1 hPa

Ambient CO

Measuring range

0 to 500 ppm

Accuracy

±5 ppm (0 to 100 ppm)

±5 % of mv (> 100 ppm)

Resolution

1 ppm

Reaction time

Approx. 35 sec

with CO probe

Ambient CO₂

Measuring range

0 to 1 Vol.%

0 to 10000 ppm

Accuracy

±50 ppm or ±2 % of mv (0 to 5000 ppm)

±100 ppm or ±3 % of mv (5001 to 10000 ppm)

Reaction time

Approx. 35 sec

with ambient CO₂ probe

Gas leak measurement for combustible gases (via gas leak detection probe)

Measuring range

0 to 10000 ppm CH₄ / C₃H₈; Display range

Accuracy

Signal optical display (LED) audible signal via buzzer

Reaction time t₉₀

< 2 sec

with gas leak detection probe

Temperature (via the fine pressure probe)

Measuring range

-40° to 2192 °F / -40 to +1200 °C max. (dependent on probe)

Accuracy

±1 °F (-40° to 212 °F) / ±0.5 °C (-40 to 100 °C)

±0.5 % of mv (Remaining Range) plus probe accuracy

Resolution

0.1 °F / 0.1 °C

General technical data

Dimensions

11 x 4 x 3 in. / 270 x 90 x 65 mm

Operating temperature

23° to 113 °F / -5 to +45 °C

Display size

240 x 320 pixels

Display function

Colour graphic display

Power supply

Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A

Maximum

500,000 readings

Storage temperature

-4° to 122 °F / -20 to +50 °C

Weight

21 oz. (without rechargeable battery) / 600 g (without rechargeable battery)

Probes

Connection cable

Order Number: 0430 0143

Mini ambient air probe

Order Number: 0600 3692

Temperature

Measuring range

32° to 176 °F / 0 to +80 °C

Temperature

Measuring range

32.0° to 212.0 °F / 0 to +100 °C

Temperature

Measuring range

32.0° to 212.0 °F / 0 to +100 °C

Resolution

0.1 °F (33.8° to 212 {#0.0} °F) / 0.1 °C (1 to 100 {+#0} °C)

Reaction time

30 sec

Ambient CO₂

Measuring range

0 to +1 Vol.%

0 to +10000 ppm

Accuracy

±(75 ppm + 3 % of mv) (0 to +5000 ppm)

±(150 ppm + 5 % of mv) (+5001 to +10000 ppm)

Gas leak probe

Order Number: 0632 3330

Ambient CO probe

Order Number: 0632 3331

Ambient CO

Measuring range

0 to +500 ppm

Accuracy

±5 ppm (0 to +100 ppm)

±5 % of mv (+100.1 to +500 ppm)

Fine pressure probe

Order Number: 0638 0330

Temperature - Type K TC

Measuring range

Depends on probe, max. -40.0° to 2192.0 °F / Depends on probe, max. -40 to +1200 °C

-328.0° to 572.0 °F (with surface probe 0604 0994 (accessory)) / -200 to +300 °C (with surface probe 0604 0994 (accessory))

Accuracy

±0.9 °F (-40° to 212.0 °F) / ±0.5 °C (-40 to +100 °C)

±0.5 % of mv (Remaining Range) additionally ±1 digit, additionally accuracy of probe

Resolution

0.1 °F / 0.1 °C

Measuring rate

max. 1 / sec

Differential Pressure

Measuring range

-40.0 to +40.0 InH₂O / -9999.9 to +9999.9 Pa

Accuracy

±0.0 InH₂O (0 to +0.04 InH₂O) / ±0.3 Pa (0 to +9.99 Pa)

±3 % of mv (+0.04 to +40.0 InH₂O) additionally ±1 digit / ±3 % of mv (+10.00 to +9999.9 Pa) additionally ±1 digit

Resolution

0.01 InH₂O (-0.60° to 0.60 InH₂O) / 0.01 Pa (-149.99 to +149.99 Pa)

0.1 InH₂O (Remaining Range) / 0.1 Pa (Remaining Range)

Measuring rate

max. 1 / sec

Velocity / Volume flow

Measuring range

Display range: 29.50 to 591 fpm / Display range: 0.15 to 3 m/s

Resolution

0.1 fpm / 0.1 m/s

Instruments / options

Fine pressure probe

Order Number: 0638 0330

Measurement cells

Product Brochures

Manuals

Software