testo 184 T2 - Temperature data logger for transport monitoring

Order-Nr.  0572 1842 01

  • Operating time 150 days

  • All in one temperature data logger: PDF measurement data, configuration file, acceptance test certificate 3.1 according to DIN ISO 10204 and instruction manual

  • Immediate creation of a PDF report directly after connection to a computer

  • Comes in packs of 10

With the temperature data logger testo 184 T2, you can now monitor the transport of sensitive products such as pharmaceuticals, foods or electronics safely and conveniently. Thanks to a start and stop button, the temperature data logger guarantees easiest data recording operation. At their destination, a glance at the display or the LEDs tells you whether the configured limit values have been adhered to during transport.

Product Description

With the temperature data loggers testo 184 T2, you monitor the transport of sensitive goods such as pharmaceuticals, food or electronics, over a maximum period of 150 days.

At their destination, a glance at the display or the LEDs tells you whether the configured limit values have been adhered to. In order to obtain detailed information, it is sufficient to connect the logger to a PC – a PDF report is immediately generated with all relevant data. In order for you to be able to work even more efficiently and conveniently with the data loggers, all required files and information are stored directly and securely in the testo 184 T2: Configuration file, calibration certificate, instruction manual and PDF report of your recorded data.

The measurement value store of the temperature data logger testo 184 T2 has a capacity of 40,000 measurement values. The measurement rate is freely selectable from 1 min to 24 h.

testo 184 T1 is compliant with GxP, 21 CFR Part 11, HACCP and EN 12830. Testo AG is an ISO 9001:2008-certified company, and ensures the adherence to the norms through internal audits as well as accredited external audits.

All advantages at a glance:
1. Clear alarm indication
A glance at the the LEDs is enough to know whether limit values have been violated during transport.
2. Easiest operation
testo 184 T2 is intuitively operated, and can be used without special training or previous knowledge: The "Start" button begins data recording. "Stop" ends it.
3. Easiest configurationIn each testo 184 T2, a configuration file is stored, with which the configuration of the testo 184 T2 is childsplay – no download, no installation, no user interface and no additional costs.
4. Convenient readout
A PDF report with the transport data is created immediately on connection of the testo 184 T2 data logger to the USB interface of a computer. This is suitable for long-term archiving according to PDF/A standard.
5. Mobile readout/printout on site
All testo 184 T2 data loggers can be read out on site with an NFC-capable Android smartphone. Data transfer from the data logger to a compatible Testo fast printer (option) also functions wirelessly via NFC

 

Delivery Scope

Temperature data logger testo 184 T2 for transport monitoring, including acceptance test certificate 3.1 according to DIN ISO 10204 (stored in testo 184 T2 ), battery and 1 piece double-sided adhesive tape for easy attachment of the temperature data logger.
Temperature - NTC

Measuring range

-35 to +70 °C

Accuracy

±0.5 °C

Resolution

0.1 °C

General technical data

Dimensions

44 x 12 x 97 mm

Operating temperature

-35 to +70 °C

Protection class

IP67 with TopSafe

Standards

EN 12830

Certificate type

Certified by HACCP International

Operating time

150 days

Alarm function

via LED and Display

Battery type

Lithium battery non exchangeable

Memory

40,000 measuring values

Storage temperature

-55 to +70 °C

Weight

45 g

Monitoring and documentation of temperature, humidity and shock in pharmaceutical logistics

Most pharmaceuticals must be continuously transported and stored at defined upper and lower temperature and humidity limit values, along the entire supply chain. A violation of these predefined upper and lower limit values can lead to irreversible alterations of the active ingredients or the composition of the medicaments.

However, not only medicaments and their active ingredients are endangered. Minus temperatures or extreme temperature fluctuations can also cause damage to containers or medical instruments. The consequence is the threat of spoilage of the entire goods, financial losses and damage to reputation.

In GxP-compliant quality management, the control of the entire cold chain, from production to delivery, is therefore a prerequisite, in order not only to ensure product quality, but also to prevent financial losses and damage to your reputation in the pharmaceutical industry.

With the testo 184 data loggers, your products can be protected from unnoticed gaps in the cold chain, and the entire distribution path monitored and documented simply, securely and compliantly with all common standards, guidelines and regulations.

Product Brochures

Manuals