1. Home
  2. CO₂ probe head including temperature and humidity sensor
0632 1550

CO₂ probe head including temperature and humidity sensor

CO2 probe head including temperature and humidity sensor
More applications, less equipment: if you already own one of our probes with a universal handle, you can easily expand your portfolio of probes with the CO₂ probe head.
  • Intuitive: parallel determination of CO₂ concentration, humidity and air temperature in indoor areas, including long-term measurement
  • Precise: absolute pressure compensation, calculation of wet bulb temperature, dew point and absolute humidity
  • Convenient: probe head is easy to attach to the universal handle with Bluetooth or cable (available to order separately)
  • Intelligent calibration concept: only send in the probe head for calibration

Temperature - NTC

Measuring range
32° to 122 °F / 0 to +50 °C
Accuracy
±0.9 °F / ±0.5 °C
Resolution
0.1 °F / 0.1 °C

Humidity - Capacitive

Measuring range
5 to 95 %rH
Accuracy
Hysteresis: ±1.0 %rH
±3 %rH (65 to 90 %rH)
±2 %rH (35 to 65 %rH)
±5 %rH (Remaining Range)
long-term stability: ±1 %RH / year
±0.06 %rF/K (32° to +122 °F) / ±0.06 %RH/K (0 to +50 °C)
±3 %rH (10 to 35 %rH)
Resolution
0.1 %rH

Absolute Pressure

Measuring range
+280 to +440 InH₂O / +700 to +1100 hPa
Accuracy
±1.2 InH₂O / ±3.0 hPa
Resolution
0.1 InH₂O / 0.1 hPa
CO2 probe head including temperature and humidity sensor and calibration protocol.

Please do not use the probe head in condensing atmospheres. For continuous application in high-humidity ranges
    > 80% RH at ≤ 30 °C for > 12 h
    > 60% RH at > 30 °C for > 12 h
please contact Testo Service or contact us via the Testo website.
Headquarters
Testo Inc.
40 White Lake Road
07871Sparta
New Jersey
800-227-0729

© 2026