Temperature - NTC | Measuring range -40 to +150 °C |
|---|---|
Accuracy ±0.4 °C (-40 to -25.1 °C) ±0.4 °C (+75 to +99.9 °C) ±0.5 % of mv (Remaining Range) ±0.3 °C (-25 to +74.9 °C) | |
Resolution 0.1 °C | |
Temperature - TC Type K (NiCr-Ni) | Measuring range -200 to +1370 °C |
|---|---|
Accuracy ±(0.3 °C + 0.3 % of mv) | |
Resolution 0.1 °C | |
Differential pressure (internal sensor) - Piezoresistive | Measuring range -150 to +150 hPa |
|---|---|
Accuracy ±0.05 hPa (0 to 1.00 hPa) (1.01 to +150 hPa) ±0.2 hPa + 1.5 % of mv | |
Resolution 0.01 hPa | |
General technical data | Weight 250 g |
|---|---|
Dimensions 154 x 65 x 32 mm | |
Operating temperature -20 to +50 °C | |
Connectable probes 1 x digital probe with cable or 1 x Temperature NTC TUC, 1 x digital Bluetooth probe or testo Smart Probe, 1 x differential pressure (internal), 1 x Temperature TC Type K | |
Product colour black/orange | |
Battery life 12 h (typically vane measurement) | |
Battery type 3 AA mignon 1.5 V | |
Interface Bluetooth®; USB | |
Storage temperature -20 to +50 °C | |
General technical data | Storage temperature -20 to +60 °C |
|---|---|
Weight 360 g | |
Dimensions 375 x 105 x 46 mm | |
Operating temperature -5 to +50 °C | |
Probe head diameter 100 mm | |
Product colour black/orange | |
Battery type 4 x type AA batteries | |
Battery life 60 h | |
Data transfer Bluetooth® | |
Radio range 20 m | |
Temperature - NTC | Measuring range -20 to +70 °C |
|---|---|
Accuracy ±0.5 °C | |
Resolution 0.1 °C | |
Velocity / Volume flow | Measuring range 0.3 to 35 m/s |
|---|---|
Accuracy ±(0.1 m/s + 1.5 % of mv) (0.3 to 20 m/s) ±(0.2 m/s + 1.5 % of mv) (20.01 to 35 m/s) | |
Resolution 0.01 m/s | |
Temperature - NTC | Measuring range -20 to +70 °C |
|---|---|
Accuracy ±0.8 °C (-20 to 0 °C) ±0.5 °C (0 to +70 °C) | |
Resolution 0.1 °C | |
Humidity - Capacitive | Measuring range 5 to 95 %RH |
|---|---|
Accuracy Hysteresis: ±1.0 %RH ±3 %RH (10 to 35 %RH) ±0.06 %RH/K (k=1) long-term stability: ±1%RH / year ±3 %RH (65 to 90 %RH) ±5 %RH (Remaining Range) ±2 %RH (35 to 65 %RH) | |
Resolution 0.1 %RH | |
Absolute Pressure | Measuring range +700 to +1100 hPa |
|---|---|
Accuracy ±3.0 hPa | |
Resolution 0.1 hPa | |
Velocity - Hot wire | Measuring range 0 to 30 m/s |
|---|---|
Accuracy ±(0.5 m/s + 5 % of mv) (0 to 20 m/s) ±(0.03 m/s + 4 % of mv) (20.01 to 30 m/s) | |
Resolution 0.01 m/s | |
General technical data | Storage temperature -20 to +70 °C |
|---|---|
Weight 35 g | |
Dimensions 235 x 9 x 9 mm | |
Operating temperature -20 to +70 °C | |
Diameter probe shaft 9 mm | |
Length probe shaft 230 mm | |
Product colour Black | |
General technical data | Weight 155 g |
|---|---|
Dimensions angle: 65 x 65 x 15 mm telescope: 375 x 17 x 16 mm | |
Operating temperature -5 to +50 °C | |
Length telescope 1,000 mm | |
Diameter telescope 12 mm | |
Product colour Black | |
Storage temperature -20 to +60 °C | |
General technical data | Weight 1400 g |
|---|---|
Dimensions 516 x 135 x 256 mm | |
Product colour Black | |