Precision pressure dew point probe for measurements in compressed air systems incl. cert. with test point -40°C tpd
Fast and accurate determination of the pressure dew point
High-precision trace humidity measurement
Pressure dew point measuring range: -40 to +50 °C tpd
The pressure dew point is an important criterion for ensuring that compressed air systems work perfectly. Use the precision pressure dew point probe with measuring chamber (combined with the appropriate measuring instrument) to determine the pressure dew point quickly and accurately.
Connection to the compressed air system is via a standard plug-in connection (G1/4” internal thread as per ISO 228-1) or via a test connection screw fitting (for measurements on vehicles). If necessary, other standard plug-in connections with a G 1/4” thread can also be screwed into the measuring chamber. The probe can also be used for indoor air humidity and temperature measurement without a measuring chamber.
Humidity - Capacitive
Humidity - Capacitive
Measuring range
0 to +100 %rH
Pressure dewpoint
Pressure dewpoint
Measuring range
-40.0° to 50.0 °Ctpd
Accuracy
±2.0 °Ctpd (-19.9 to -10.0 °Ctpd)
±3.0 °Ctpd (-29.9 to -20.0 °Ctpd)
±0.8 °Ctpd (-4.9 to 50.0 °Ctpd)
±4.0 °Ctpd (-40.0 to -30.0 °Ctpd)
±1.0 °Ctpd (-9.9 to -5.0 °Ctpd)
General technical data
General technical data
Length probe shaft
11.8 in. / 300 mm
Precision pressure dew point probe with measuring chamber, including certificate with -40 °C tpd. Note: You need a plug-in head cable (0430 0143) for this probe.