Weight | 595 g |
---|---|
Dimensions | 77 x 109 x 63 mm |
Operating temperature | -10 to +50 °C |
Protection class | IP54 |
System requirements | requires iOS 12.0 or newer; requires Android 6.0 or newer; requires mobile end device with Bluetooth 4.0 |
Product colour | Black |
Auto-off | 10 |
Battery type | 3 AAA micro batteries |
Battery life | 130 h |
Data transfer | Bluetooth® |
Radio range | 150 m |
Refrigerants in instrument | R407H; R23; R290; R1234yf; R442A; R134a; R123; R452B; R438A; R421A; R453a; R401B; R1233zd; R410A; R407A; R407C; R452A; R401A; R422C; R427A; R422B; R424A; R125; R414B; R404A; R408A; R448A; R422D; R12; R114; R421B; R449A; R409A; R420A; R444B; R32; R454A; R124; R402A; R450A; R407F; R416A; R13; R437A; R1234ze; R454B; R402B; R22; R434A; R454C; R455A; R458A; R500; R502; R503; R507; R513A; R600a; R718 (Hâ‚‚O); R744 (COâ‚‚) |
Refrigerants update via App | R11; FX80; I12A; R1150; R1270; R13B1; R14; R142B; R152A; R161; R170; R227; R236fa; R245fa; R401C; R406A; R407B; R407D; R41; R411A; R412A; R413A; R417A; R417B; R417C; R422A; R426A; R508A; R508B; R600; RIS89; SP22 |
Storage temperature | -20 to +60 °C |
* when not connected via Bluetooth |
Measuring range | -1 to 60 bar |
---|---|
Accuracy | ±0.5 % fs |
Resolution | 0.01 bar |
Probe connection | 3 x 7/16" – UNF |
Overload rel. (high pressure) | 65 bar |
Measuring range | -40 to +150 °C |
---|---|
Accuracy | ±1.3 °C (-20 to +85 °C) |
Resolution | 0.1 °C |
Weight | 127.4 g |
---|---|
Dimensions | 183 x 90 x 30 mm |
Operating temperature | -20 to +50 °C |
Product-/housing material | Plastic |
System requirements | requires iOS 13.0 or newer; requires Android 8.0 or newer; requires mobile end device with Bluetooth 4.0 |
Product colour | black/orange |
Battery type | 3 AAA micro batteries |
Battery life | 150 h |
Data transfer | Bluetooth® |
Radio range | 100 m |
Storage temperature | -20 to +60 °C |
Measuring range | -20 to +500 °C |
---|---|
Accuracy | ±1 °C (0 to +100 °C) |
Resolution | 0.2 °C |
Measuring range | 1.0 to 600.0 V |
---|---|
Resolution | max. 1 mV |
Accuracy | ± (0.8 % of mv + 3 Digit) |
Measuring range | 1.0 mV to 600.0 V |
---|---|
Resolution | max. 1 mV |
Accuracy | ± (1 % of mv + 3 Digit) |
Measuring range | 0.1 to 600.0 A (clamp) 0.1 to 600.0 μA (measuring cables) |
---|---|
Resolution | 0.1 (clamp) 0.1 (measuring cables) |
Accuracy | ± (2 % of mv + 5 Digit) (measuring cables) ± (2 % of mv + 5 Digit) (clamp) |
Measuring range | 0.1 to 600.0 A (clamp) 0.1 to 600.0 μA (measuring cables) |
---|---|
Resolution | 0.1 A (clamp) 0.1 μA (measuring cables) |
Accuracy | ± (2 % of mv + 5 Digit) (clamp) ± (1.5 % of mv + 5 Digit) (measuring cables) |
Resolution | max. 0.01 Ohm |
---|---|
Measuring range | 0.01 to 60.0 MOhm |
Accuracy | ± (1.5 % of mv + 3 Digit) |
Resolution | max. 0.01 nF |
---|---|
Measuring range | 2 nF to 60.00 mF |
Accuracy | ± (1.5 % of mv + 5 Digit) |
Resolution | max. 0.1 Hz |
---|---|
Measuring range | 0.01 Hz to 9999 kHz |
Accuracy | ± (1 % of mv + 1 Digit) |
Measuring range | 0.1 W to 600 kW |
---|---|
Accuracy | ± (5 % of mv + 5 Digit) |
Resolution | 0.1 W |
Measuring range | 0.1 to 600.0 kvar |
---|---|
Accuracy | ± (10 % of mv + 5 Digit) |
Resolution | 0.1 var |
Measuring range | 0.1 VA to 600.0 kVA |
---|---|
Accuracy | ± 1 Digit |
Resolution | 0.1 VA |
Measuring range | 0.1 W to 600.0 kW |
---|---|
Accuracy | ± 1 Digit |
Resolution | max. 0.1 W |
Measuring range | -1.00 to +1.00 |
---|---|
Accuracy | ± (10 % of mv + 5 Digit) |
Resolution | 0.01 |
Operating humidity | 0 to 80.0 %RH |
---|---|
Weight | 378 g |
Dimensions | 249 x 96 x 43 mm |
Operating temperature | -10 to +50 °C |
Connectable probes | 1 x temperature probe (adapter required for type K thermocouples) |
System requirements | requires iOS 11.0 or newer; requires Android 6.0 or newer; requires mobile end device with Bluetooth 4.0 |
Product colour | Black |
Standards | EN 61326-1; EN 61010-2-032; EN 61140 |
Battery type | 3 AAA micro batteries |
Display type | LCD (Liquid Crystal Display) |
Display size | 2 lines |
Interface | 2 measuring jacks |
Storage temperature | -15 to +60 °C |
Overvoltage Category | CAT IV 600V; CAT III 1000V |
Authorizations | CSA; CE |
Internal_resistance | 11 MOhm (AC / DC) |
Diode test | yes |
---|---|
Continuity testing | yes |
Display ligthing | yes |
Hold | yes |
MinMax | yes |
Power Measurement | yes |
Temperature measurement with adapter | yes |
True RMS | yes |
Inrush current measurement | yes |
Bluetooth | yes |
Display (counts) | 6,000 |
μA current measuring range | yes |
Autom. measurement parameter detection | yes |
System requirements | requires iOS 13.0 or newer; requires Android 8.0 or newer; requires mobile end device with Bluetooth 4.0 |
---|---|